10 Spectrum Emission Mask Measurement
Source
Query ARB Memory File List (Remote Command Only)
Queries the test set for the list of waveform segments in the ARB memory.
This command returns a string for waveform segment names in ARB memory. If you want a string list of
waveform segments in the ARB memory, use
"Query ARB Memory Full File List (Remote Command Only)"
Remote Command
:SOURce:RADio:ARB:CATalog?
Example
:SOUR:RAD:ARB:CATalog?
Notes
The return data is in the following format:
<integer> - memory used
<integer> - memory free
<string> … - comma separated list of waveform segments within ARB memory
Initial S/W Revision
A.05.00
Query ARB Memory Full File List (Remote Command Only)
Queries the test set for the string list of waveform segments in the ARB memory. It returns a string list for
waveform segment names in the ARB memory.
Remote Command
:SOURce:RADio:ARB:FCATalog?
Example
:SOUR:RAD:ARB:FCATalog?
Notes
The return data is in the following format:
<integer> - memory used
<integer> - memory free
<integer> - file count in ARB memory
<string>,<string>, … <string> - comma separated string list of waveform segments within ARB
memory
Example: SOUR:RAD:ARB:FCAT?
EXT returns: 27499,2069653,3,”c2k.wfm”,”gsm.wfm”,”wcdma.wfm”
Initial S/W Revision
A.09.00
ARB Setup
Allows access to the ARB setup sub-menus.
Key Path
Source, Modulation Setup, ARB
Initial S/W Revision
A.05.00
Sample Rate
Allows you to set the ARB waveform playback sample rate.
W-CDMA & HSPA Measurement Applications Reference
1177
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...