![Keysight E6640A Measurement Applications User & Programmer Reference Download Page 220](http://html1.mh-extra.com/html/keysight/e6640a/e6640a_measurement-applications-user-and-programmer-reference_1961645220.webp)
4 Input/Output Functions
Input/Output
Key Path
Input/Output
Preset
All settings under this key are returned to their default state when Restore Input/Output Defaults is
pressed.
State Saved
Saved in State
Initial S/W Revision
A.09.49
RF Output
Specifies the RF Output Port used.
Switching from the RF Output port to one of the RFIO ports changes the transmitter performance of the
instrument.
Key Path
Input/Output, RF Output & Test Set Config
Remote Command
[:SENSe]:FEED:RF:PORT:OUTPut RFOut | RFIO1 | RFIO2 | GPSout | GNSSout |
RFIO3 | RFIO4
[:SENSe]:FEED:RF:PORT:OUTPut?
Example
:FEED:RF:PORT:OUTP RFIO1
Preset
This is unaffected by Mode Preset but is set to RFOut on a "Restore Input/Output Defaults" or
"Restore System Defaults -> All"
State Saved
Saved in State
Readback
The current RF Output Port selected is read back to this key
Initial S/W Revision
A.05.01
Modified at S/W Revision
A.14.00
RF Output
The RF port that will be used for the current output.
Key Path
Input/Output, RF Output & Test Set Config, RF Output
Example
:FEED:RF:PORT:OUTP RFO
Notes
On E6640A with hardware M9430A, if RF Output is selected as RF Output Port, you need to choose
the settings in the Half Duplex Config menu to determine which port (RFIO3 or RFIO4) will be used.
On E6640A with hardware M9431A, this setting is not supported. If the SCPI command is sent with
this setting, an error is generated, –221, “Settings conflict;option not installed”
ReadBack
RF Output
Initial S/W Revision
A.05.01
Modified at S/W Revision
A.14.00
220
W-CDMA & HSPA Measurement Applications Reference
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...