10 Spectrum Emission Mask Measurement
Sweep/Control
GSM: 4.615383
State Saved
Saved in instrument state
Min
100.000 ns
Max
559.0000 ms
Default Unit
S
Initial S/W Revision
Prior to A.02.00
Offset
Adjusts the accumulated offset between the periodic timer events and the trigger event. Adjusting the
accumulated offset is different than setting an offset, and requires explanation.
The periodic timer is usually not synchronized with any external events, so the timing of its output events
has no absolute meaning. Since the timing relative to external events (RF signals) is important, you need to
be able to adjust (offset) it. However, you have no direct way to see when the periodic timer events occur.
All that you can see is the trigger timing. When you want to adjust the trigger timing, you will be changing
the internal offset between the periodic timer events and the trigger event. Because the absolute value of
that internal offset is unknown, we will just call that the accumulated offset. Whenever the Offset
parameter is changed, you are changing that accumulated offset. You can reset the displayed offset using
Reset Offset Display. Changing the display does not change the value of the accumulated offset, and you
can still make additional changes to accumulated offset.
To avoid ambiguity, we define that an increase in the "offset" parameter, either from the knob or the SCPI
adjust command, serves to delay the timing of the trigger event.
Key Path
Trigger, Periodic Timer
Remote Command
:TRIGger[:SEQuence]:FRAMe:OFFSet <time>
:TRIGger[:SEQuence]:FRAMe:OFFSet?
Example
TRIG:FRAM:OFFS 1.2 ms
Notes
The front panel interface (for example, the knob), and this command, adjust the accumulated offset,
which is shown on the active function display. However, the actual amount sent to the hardware
each time the offset is updated is the delta value, that is, the current accumulated offset value minus
the previous accumulated offset value. Note that the accumulated offset value is essentially arbitrary;
it represents the accumulated offset from the last time the offset was zeroed (with the Reset Offset
Display key).
Note that this command does not change the period of the trigger waveform. Note also that Offset is
used only when the sync source is set to OFF, otherwise delay is used, see section
An increase in the "offset" parameter, either from the knob or the SCPI adjust command, serves to
delay the timing of the trigger event.
Notes
When the SCPI command is sent the value shown on the key (and the Active Function, if this happens
to be the active function) is updated with the new value. However, the actual amount sent to the
hardware is the delta value, that is, the current accumulated offset value minus the previous
accumulated offset value.
The SCPI query simply returns the value currently showing on the key.
W-CDMA & HSPA Measurement Applications Reference
1283
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...