15 Power Control Measurement
Meas
Capture Interval (Remote Command only)
Enables you to specify the capture interval in frame units.
Mode
WCDMA
Remote Command
[:SENSe]:PCONtrol:CAPTure:TIME <real>
[:SENSe]:PCONtrol:CAPTure:TIME?
Example
PCON:CAPT:TIME 4.0
PCON:CAPT:TIME?
Notes
For consistency, this parameter must be controlled with a float numeric from remote commands, but
the front panel interface uses Enum for easier operation.
You must be in the W-CDMA mode to use this command. Use INSTrument:SELect to set the mode.
Dependencies
This Parameter is coupled with MeasType. 8 Frames cannot be set (menu key is grayed out) when
MeasType is SlotPhase.
Couplings
See Dependencies
Preset
4.0
State Saved
Saved in instrument state.
Range
1.0|2.0|4.0|8.0
Initial S/W Revision
Prior to A.02.00
Scramble Code
Sets the mobile station (MS) scramble code for synchronization. When this key is selected, a Hexadecimal
Input menu appears.
Key Path
Meas Setup
Mode
WCDMA
Remote Command
[:SENSe]:PCONtrol:SYNC:SCRamble <integer>
[:SENSe]:PCONtrol:SYNC:SCRamble?
Example
PCON:SYNC:SCR 0
PCON:SYNC:SCR?
Preset
0 (0x0)
State Saved
Saved in instrument state.
Min
0 (0x0; 24 bits)
Max
16777215 (0xFFFFFF; 24 bits)
Initial S/W Revision
Prior to A.02.00
Meas Preset
Restores all the measurement parameters to their default values.
2674
W-CDMA & HSPA Measurement Applications Reference
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...