9 ACP Measurement
FREQ Channel
Key Path
FREQ Channel
Remote Command
[:SENSe]:FREQuency:CENTer:STEP[:INCRement] <freq>
[:SENSe]:FREQuency:CENTer:STEP[:INCRement]?
[:SENSe]:FREQuency:CENTer:STEP:AUTO OFF|ON|0|1
[:SENSe]:FREQuency:CENTer:STEP:AUTO?
Example
FREQ:CENT:STEP:AUTO ON
FREQ:CENT:STEP 500 MHz
FREQ:CENT UP increases the current center frequency value by 500 MHz
FREQ:CENT:STEP?
FREQ:CENT:STEP:AUTO?
Notes
Preset and Max values are depending on Hardware Options (503, 507, 508, 513, 526)
Notes
Preset and Max values are dependent on Hardware Options (5xx)
Dependencies
Freq Offset is not available in External Mixing. In this case the Freq Offset key is grayed out and shows
a value of zero. It will once again be available, and show the previously set value, when you return to
the RF Input.
Dependencies
Span, RBW, Center frequency
If the electronic/soft attenuator is enabled, any attempt to change the value of the center frequency
>3.6 GHz by pressing the Up-arrow key, fails and results in an advisory message. If the equivalent
SCPI command is sent, this same message is generated as part of a “–221, Settings conflict”
warning.
Couplings
When auto-coupled in a non-zero span, the center frequency step size is set to 10% of the span.
When auto-coupled in zero span, the center frequency step size is set to the equivalent –3 dB RBW
value.
Preset
Auto
ADEMOD: 1 MHz
ON
State Saved
Saved in instrument state
Min
– (the maximum frequency of the instrument). That is, 27 GHz max freq instrument has a CF step
range of +/– 27 GHz. Note that this is the maximum frequency given the current settings of the
instrument, so in External Mixing, for example, it is the maximum frequency of the current mixer
band.
Max
The maximum frequency of the instrument. That is, 27 GHz max freq instrument has a CF step range
of +/– 27 GHz. Note that this is the maximum frequency given the current settings of the instrument,
so in External Mixing, for example, it is the maximum frequency of the current mixer band.
Default Unit
Hz
Status Bits/OPC
dependencies
non-overlapped
Initial S/W Revision
Prior to A.02.00
Modified at S/W Revision
A.03.00
W-CDMA & HSPA Measurement Applications Reference
679
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...