![Keysight E6640A Measurement Applications User & Programmer Reference Download Page 1591](http://html1.mh-extra.com/html/keysight/e6640a/e6640a_measurement-applications-user-and-programmer-reference_19616451591.webp)
11 Occupied Bandwidth Measurement
Sweep/Control
Key Path
Trigger, External 2
Remote Command
:TRIGger[:SEQuence]:EXTernal2:LEVel
:TRIGger[:SEQuence]:EXTernal2:LEVel?
Example
TRIG:EXT2:LEV 1.1 V
Couplings
This same level is used for the Ext2 trigger source in the Trigger menu, for the Ext2 selection in the
Periodic Timer sync source (in the Trigger menu and in the Gate Source menu), and also for the Ext2
selection in the Gate Source menu.
Preset
1.2 V
State Saved
Saved in instrument state
Min
–5 V
Max
5 V
Default Unit
V
Backwards
Compatibility SCPI
:TRIGger[:SEQuence]:FRAMe:EXTernal2:LEVel
Initial S/W Revision
Prior to A.02.00
Trig Slope
Controls the trigger polarity. It is set positive to trigger on a rising edge and negative to trigger on a falling
edge.
Key Path
Trigger, External 2
Remote Command
:TRIGger[:SEQuence]:EXTernal2:SLOPe POSitive|NEGative
:TRIGger[:SEQuence]:EXTernal2:SLOPe?
Example
TRIG:EXT2:SLOP NEG
Couplings
This same slope is used in the Ext2 selection for the trigger source in the Trigger menu and for the
period timer sync source (in the Trigger menu and in the Gate Source menu).
Preset
POSitive
State Saved
Saved in instrument state
Backwards
Compatibility SCPI
:TRIGger[:SEQuence]:FRAMe:EXTernal2:SLOPe
Backwards Compatibility
Notes
The legacy :TRIGger[:SEQuence]:SLOPe command affects the slopes for the VID, LINE, EXT1, EXT2,
and RFB triggers.
Initial S/W Revision
Prior to A.02.00
RF Burst
Pressing this key, when it is not selected, selects the RF Burst as the trigger. A new sweep/measurement
will start when an RF burst envelope signal is identified from the signal at the RF Input connector. Pressing
this key, when it is already selected, accesses the RF Burst trigger setup menu.
W-CDMA & HSPA Measurement Applications Reference
1591
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...