8 Channel Power Measurement
Source
Key Path
Source, Modulation Setup, ARB, ARB Setup
Remote Command
:SOURce:RADio:ARB:BASeband:FREQuency:OFFSet <freq>
:SOURce:RADio:ARB:BASeband:FREQuency:OFFSet?
Example
:SOUR:RAD:ARB:BAS:FREQ:OFFS 0.00 Hz
Dependencies
When a new waveform is selected for playback the settings contained within the associated
waveform header file are applied to the ARB. The baseband frequency offset is one of the values
stored within the header file. If the newly selected waveform file has an associated header file, the
baseband frequency offset is updated with the value from the header file. The baseband frequency
offset will remain unchanged if the newly selected waveform does not have an associated header file.
Preset
0.00 Hz
Min
–50.00 MHz
Max
50.00 MHz
Initial S/W Revision
A.05.00
Trigger Type
Allows access to the trigger type sub-menus. The setting for trigger type determines the behavior of the
waveform when it plays.
Key Path
Source, Modulation Setup, ARB
Remote Command
:SOURce:RADio:ARB:TRIGger:TYPE CONTinuous | SINGle | SADVance GATE
:SOURce:RADio:ARB:TRIGger:TYPE?
Example
:SOUR:RAD:ARB:TRIG:TYPE CONT
:SOUR:RAD:ARB:TRIG:TYPE?
Notes
Gated trigger type will be implemented at a later release
Preset
CONTinuous
Range
Continuous | Single | Seg Adv | Gated
Initial S/W Revision
A.05.00
Continuous
Sets the active trigger type to Continuous. If Continuous is already selected as the active trigger type,
pressing this key allows access to the continuous trigger type setup menu. In Continuous trigger mode, the
waveform repeats continuously.
Key Path
Source, Modulation Setup, ARB, Trigger Type
Remote Command
:SOURce:RADio:ARB:TRIGger:TYPE:CONTinuous[:TYPE] FREE | TRIGger | RESet
:SOURce:RADio:ARB:TRIGger:TYPE:CONTinuous[:TYPE]?
Example
:SOUR:RAD:ARB:TRIG:TYPE:CONT FREE
Preset
FREE
510
W-CDMA & HSPA Measurement Applications Reference
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...