3 Programming the Test Set
STATus Subsystem
The register requires that the associated PTR or NTR filters be set before a condition register bit can set a
bit in the event register. The data in this register is latched until it is queried. Once queried, the register is
cleared
Mode
All
Remote Command
:STATus:QUEStionable:TEMPerature[:EVENt]?
Example
STAT:QUES:TEMP?
Preset
0
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Temperature Negative Transition
This command determines which bits in the Questionable Temperature Condition register will set the
corresponding bit in the Questionable Temperature Event register when the condition register bit has a
negative transition (1 to 0). The variable <integer> is the sum of the decimal values of the bits that you want
to enable.
Mode
All
Remote Command
:STATus:QUEStionable:TEMPerature:NTRansition <integer>
:STATus:QUEStionable:TEMPerature:NTRansition?
Example
STAT:QUES:TEMP:NTR 1 Reference Oscillator Oven not cold will be reported to the Temperature
Summary of the Status Questionable register.
Preset
0
Min
0
Max
32767
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Temperature Positive Transition
This command determines which bits in the Questionable Temperature Condition register will set the
corresponding bit in the Questionable Temperature Event register when the condition register bit has a
positive transition (0 to 1). The variable <integer> is the sum of the decimal values of the bits that you want
to enable.
Mode
All
Remote Command
:STATus:QUEStionable:TEMPerature:PTRansition <integer>
:STATus:QUEStionable:TEMPerature:PTRansition?
Example
STAT:QUES:TEMP:PTR 1 Reference Oscillator Oven became cold will be reported to the
W-CDMA & HSPA Measurement Applications Reference
177
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...