![Keysight E6640A Measurement Applications User & Programmer Reference Download Page 1741](http://html1.mh-extra.com/html/keysight/e6640a/e6640a_measurement-applications-user-and-programmer-reference_19616451741.webp)
12 Code Domain Measurement
Meas Setup
15000,1, QPSK
CDP:SBO:LIST:CHAN? 3
15000,3, QPSK
CDP:SBO:LIST:CHAN? 4
15000,16, QPSK
CDP:SBO:LIST:CHAN? 5
240000,15, QAM16
Notes
The maximum number of entries is 512.
Error messages associated with this parameter.
One of the following error messages is logged if the given parameter is invalid.
If an error is reported, the SCPI command is rejected and the instrument’s settings do not change.
(1) “Missing Parameter”
This error is reported if the number of parameters is less than 4.
For example,
:SENSe:CDPower:SBOundary:LIST:BTS:APPend, 15000, 0
The 3rd parameter is missing.
(2) “Illegal parameter value”
This error is reported if parameter type is invalid or if enum value is invalid.
For example,
:SENSe:CDPower:SBOundary:LIST:BTS:APPend 15000, ON, QPSK
The 2nd parameter must be an integer.
:SENSe:CDPower:SBOundary:LIST:BTS:APPend, 15001, 8, QPSK
The 1st parameter value (Symbol Rate) is not allowed.
Only the values given in Range field are valid for the Symbol Rate. You may specify these in numeric
form, but they are interpreted as an enumeration and the error results if the value does not translate
to one in the list.
(3) “Data out of range”
This error is reported if parameter value is out of range.
For example,
:SENSe:CDPower:SBOundary:LIST:BTS:APPend 15000, 256, QPSK
The 2nd parameter is out of range.
(1)“Setting Conflict”
This error is reported if the given code channel overlaps other code channel on code domain.
For example, if a user sends the following two commands, the second command causes the error
message because C7(0) overlaps C8(0).
:SENSe:CDPower:SBOundary:LIST:BTS:INIT, 15000, 0, QPSK
The example above is OK
:SENSe:CDPower:SBOundary:LIST:BTS:APPend 30000, 0, QPSK
C7(0) overlaps C8(0)
Dependencies
This command is effective when [:SENSe]:RADio:DEVice is set to BTS and
[:SENSe]:CDPower:SBOundary[:BTS] is set to CUSTom.
W-CDMA & HSPA Measurement Applications Reference
1741
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...