12 Code Domain Measurement
Meas Setup
Meas Interval
Sets the length of the measurement interval in slots (1 slot = 666.6us).
The sum of CALCulate:CDPower:SWEep:TIME and CALCulate:CDPower:SWEep:OFFSet must be equal to
or less than SENSe:CDPower:CAPTure:TIME * 15. If the sum exceeds this value,
CALCulate:CDPower:SWEep:OFFSet is adjusted automatically.
Key Path
Meas Setup
Mode
WCDMA
Remote Command
:CALCulate:CDPower:SWEep:TIME <integer>
:CALCulate:CDPower:SWEep:TIME?
Example
CALC:CDP:SWE:TIME 2
CALC:CDP:SWE:TIME?
Notes
The sum of
and
must be equal to or less
than Capture Interval in slot.
If the sum is larger than the limit value,
Preset
1
State Saved
Saved in instrument state.
Min
1
Max
Capture Interval in slot
Initial S/W Revision
Prior to A.02.00
Meas Offset
Sets the timing offset of measurement interval in slots. (1 slot = 666.6 us)
The sum of CALCulate:CDPower:SWEep:TIME and CALCulate:CDPower:SWEep:OFFSet must be equal to
or less than SENSe:CDPower:CAPTure:TIME * 15. If the sum exceeds this value,
CALCulate:CDPower:SWEep:OFFSet is adjusted automatically.
Key Path
Meas Setup
Mode
WCDMA
Remote Command
:CALCulate:CDPower:SWEep:OFFSet <integer>
:CALCulate:CDPower:SWEep:OFFSet?
Example
CALC:CDP:SWE:OFFS 2
CALC:CDP:SWE:OFFS?
Notes
The sum of Meas Offset and
must be equal to or less than Capture
Interval. If the sum becomes more than the value, the
is adjusted
automatically to Capture Interval – Meas Offset.
Preset
0
State Saved
Saved in instrument state.
1718
W-CDMA & HSPA Measurement Applications Reference
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...