9 ACP Measurement
Sweep/Control
Gate Preset (Remote Command Only)
Presets the time-gated spectrum analysis capability.
This command sets gate parameter values to the ESA preset values, as follows:
Gate trigger type = edge
Gate polarity = positive
Gate delay = 1 us
Gate length = 1 us
Remote Command
[:SENSe]:SWEep:TIME:GATE:PRESet ESA Compatibility
Initial S/W Revision
Prior to A.02.00
Gate Level (Remote Command Only)
Sets the gate input transition point level for the external TRIGGER inputs on the front and rear panel. This is
a legacy command for PSA compatibility. It is simply an alias to the equivalent trigger level command.
Remote Command
[:SENSe]:SWEep:EGATe:EXTernal[1]|2:LEVel <voltage>
[:SENSe]:SWEep:EGATe:EXTernal[1]|2:LEVel?
Notes
This command is simply an alias to
:TRIGger[:SEQuence]:EXTernal[1]|2:LEVel
For details refer
Initial S/W Revision
Prior to A.02.00
Gate Polarity (Remote Command Only)
Sets the polarity for the gate signal. This setup is now done using the gate trigger’s slope setting.
When Positive (Pos) is selected, a positive-going edge (Edge) or a high voltage (Level) will satisfy the gate
condition, after the delay set with the Gate Delay key. When Negative (Neg) is selected, a negative-going
edge (Edge) or a low voltage (Level) will satisfy the gate condition after the delay.
Remote Command
[:SENSe]:SWEep:EGATe:POLarity NEGative|POSitive
[:SENSe]:SWEep:EGATe:POLarity?
Example
SWE:EGAT:POL NEG
SWE:EGAT:POL?
Preset
POSitive
State Saved
Saved in instrument state
Backwards
Compatibility SCPI
[:SENSe]:SWEep:TIME:GATE:POLarity ESA compatibility
Initial S/W Revision
Prior to A.02.00
958
W-CDMA & HSPA Measurement Applications Reference
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...