![Keysight E6640A Measurement Applications User & Programmer Reference Download Page 181](http://html1.mh-extra.com/html/keysight/e6640a/e6640a_measurement-applications-user-and-programmer-reference_1961645181.webp)
3 Programming the Test Set
Common Commands
Mode
All
Remote Command
*CAL?
Example
*CAL?
Notes
*CAL? returns 0 if successful
*CAL? returns 1 if failed
:CALibration[:ALL]? is the same as *CAL?
See additional remarks described with :CALibration[:ALL]?
Everything about :CALibration[:ALL]? is synonymous with *CAL? including all conditions, status
register bits, and couplings
Initial S/W Revision
Prior to A.02.00
Mode
All
Remote Command
:CALibration[:ALL]:NPENding
Example
CAL:NPEN
Notes
:CALibration[:ALL]:NPENding is the same as :CALibration[:ALL] including all conditions, status
register bits, except this scpi command does not BLOCK the scpi session, so the user should use
status register bits to query if the calibration is successfully completed or not.
Typical usage is:
1) :CALibration:ALL:NPENding (Start a calibration)
2) :STATus:OPERation:CONDition? (Check if the calibration is completed or not, If bit 0 is set, then
the system is doing calibration, the user should repeat this scpi query until the bit is cleared )
3):STATus:QUEStionable:CALibration:CONDition? (Check if if there are any errors/failures in previous
calibration procedure
Initial S/W Revision
X.14.20
Clear Status
Clears the status byte register. It does this by emptying the error queue and clearing all bits in all of the
event registers. The status byte register summarizes the states of the other registers. It is also responsible
for generating service requests.
Key Path
No equivalent key. Related key System, Show Errors, Clear Error Queue
Remote Command
*CLS
Example
*CLS Clears the error queue and the Status Byte Register.
Notes
For related commands, see the SYSTem:ERRor[:NEXT]? command. See also the STATus:PRESet
command and all commands in the STATus subsystem.
Status Bits/OPC
dependencies
Resets all bits in all event registers to 0, which resets all the status byte register bits to 0 also.
Backwards Compatibility
Notes
In general the status bits used in the X-Series status system will be backwards compatible with ESA
and PSA. However, note that all conditions will generate events that go into the event log, and some
W-CDMA & HSPA Measurement Applications Reference
181
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...