12 Code Domain Measurement
Meas
Trigger Source
Example
CALC:FPOW:POW1:DEF "TriggerSource=Ext1"
Notes
The trigger source parameter allows the user to choose between measurement’s triggering freely or
controlled by an external input. Ext1 and Ext2 correspond to Trigger 1 In and Trigger 2 In, respectively.
Preset
Free
Range
Free, Ext1, Ext2
Initial S/W
Revision
A.14.00
Trigger Timeout
Example
CALC:FPOW:POW1:DEF "TriggerTimeout=0.1"
Notes
The trigger timeout parameter sets the time in which the analyzer will wait for a trigger before
automatically performing the measurement.
Preset
1 s
Range
0 – 1 s
Default Unit
Seconds
Initial S/W
Revision
A.14.00
Signal Input
Example
CALC:FPOW:POW1:DEF "SignalInput=Fp50MHzCW"
Notes
The signal input parameter allows the user to select between using the main RF input or the internal
analyzer reference CW signal of 50 MHz.
Preset
FpMainRf
Range
FpMainRf, Fp50MHzCW
Initial S/W
Revision
A.14.00
Use Preselector
Example
CALC:FPOW:POW1:DEF "UsePreSelector=True"
Notes
The preselector parameter allows the user to either utilize or bypass the front end tunable filter at
frequencies above 3.6 GHz. For frequencies below 3.6 GHz, the preselector is automatically
bypassed, so you do not need to set this parameter to False in those cases.
Preset
False
Range
True (use preselector above 3.6 GHz), or False (preselector bypassed)
1706
W-CDMA & HSPA Measurement Applications Reference
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...