![Keysight E6640A Measurement Applications User & Programmer Reference Download Page 801](http://html1.mh-extra.com/html/keysight/e6640a/e6640a_measurement-applications-user-and-programmer-reference_1961645801.webp)
9 ACP Measurement
Source
RF Power
Allows you to adjust the power level of the source using the numeric keypad, step keys, or RPG. Pressing
any digit, 0 through 9, on the numeric keypad brings up the unit terminator.
Please refer to the
table below for the valid ranges.
Key Path
Source, Amplitude
Remote Command
:SOURce:POWer[:LEVel][:IMMediate][:AMPLitude] <ampl>
:SOURce:POWer[:LEVel][:IMMediate][:AMPLitude]?
Example
:SOUR:POW –100 dBm
Notes
Amplitude corrections can be specified for use with the source. In the event of amplitude corrections
being applied, the valid ranges for the RF power do not change dependant on the current amplitude
correction setting. If the combination of RF power + amplitude correction is higher or lower than the
source output range, the Source Unleveled bit is set and the “Source Unleveled” indicator will appear
on status panel to indicate that the source cannot maintain the output power that has been
requested.
When signal generator is unable to maintain the requested output level, the “Source Unleveled”
indicator will appear on status panel. When the source output setting is restored to the normal range,
the “Source Unleveled” is removed from status panel.
Internal source has list sequence mode, which comprises of several steps which contain separate
output power, frequency and waveform etc. When the source list sequence playing is complete, the
last step keeps playing, and user can use this command to change the list sequence last step’s output
power.
The multiport adapter RFIO TX ports and GPS ports cannot ensure power accuracy when power
setting is lower than –130dBm, this power setting value is defined by the sum of RF Power setting
and related amplitude correction value. But user settable value could be lower than this limit. When
application detected there exists power setting lower than –130dBm on MPA RFIO TX ports, then
popup warning message . When application detected there exists power setting lower than –
130dBm on MPA GPS ports, then popup warning message . This is only warning message, and check
is performed when RF is ON.
Notes
The Min and Max value here defined UI settable amplitude range. This range is larger than actual
amplitude range with level accuracy defined in spec.
Dependencies
The RF power is dependent on the RF output port and frequency, such that the current frequency and
selected output port determine the valid range of power values.
Preset
–100 dBm
Min
The range of values depends on the current frequency and selected RF output port. Please refer to the
table below for the valid ranges.
Max
The range of values depends on the current frequency and selected RF output port. Please refer to the
table below for the valid ranges.
Initial S/W Revision
A.05.00
W-CDMA & HSPA Measurement Applications Reference
801
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...