![Keysight E6640A Measurement Applications User & Programmer Reference Download Page 1644](http://html1.mh-extra.com/html/keysight/e6640a/e6640a_measurement-applications-user-and-programmer-reference_19616451644.webp)
12 Code Domain Measurement
AMPTD Y Scale
I/Q Error View Phase Error Window Y Ref Value
Sets the absolute power reference value for the phase error view in the I/Q Error window.
Key Path
AMPTD Y Scale
Mode
WCDMA
Remote Command
:DISPlay:CDPower:VIEW3:WINDow2:TRACe:Y[:SCALe]:RLEVel <real>
:DISPlay:CDPower:VIEW3:WINDow2:TRACe:Y[:SCALe]:RLEVel?
Example
DISP:CDP:VIEW3:WIND2:TRAC:Y:RLEV 1.0
DISP:CDP:VIEW3:WIND2:TRAC:Y:RLEV?
Notes
Target window to control depends on the SubOpCode.
Couplings
When
is set to On, this value is automatically determined by the
measurement result. When you set this value manually, Auto Scaling automatically changes to Off.
Preset
0.00
State Saved
Saved in instrument state.
Min
–360
Max
360
Initial S/W Revision
Prior to A.02.00
I/Q Error View EVM Window Y Ref Value
Sets the absolute power reference value for the EVM view in the I/Q Error window.
Key Path
AMPTD Y Scale
Mode
WCDMA
Remote Command
:DISPlay:CDPower:VIEW3:WINDow3:TRACe:Y[:SCALe]:RLEVel <real>
:DISPlay:CDPower:VIEW3:WINDow3:TRACe:Y[:SCALe]:RLEVel?
Example
DISP:CDP:VIEW3:WIND3:TRAC:Y:RLEV 1.0
DISP:CDP:VIEW3:WIND3:TRAC:Y:RLEV?
Notes
Target window to control depends on the SubOpCode.
Couplings
When
is set to On, this value is automatically determined by the
measurement result. When you set this value manually, Auto Scaling automatically changes to Off.
Preset
0.00
State Saved
Saved in instrument state.
Min
–100
Max
100
Initial S/W Revision
Prior to A.02.00
Code Domain View CDP Window Y Ref Value
Sets the absolute power reference value for the code domain power view in the CDP window.
1644
W-CDMA & HSPA Measurement Applications Reference
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...