![Keysight E6640A Measurement Applications User & Programmer Reference Download Page 2658](http://html1.mh-extra.com/html/keysight/e6640a/e6640a_measurement-applications-user-and-programmer-reference_19616452658.webp)
15 Power Control Measurement
Meas
Preset
0 Hz
Default Unit
Hz
Initial S/W
Revision
A.14.00
Trigger Delay
Example
CALC:FPOW:POW1:DEF "TriggerDelay=0.025"
Notes
The trigger delay parameter sets the time after an external trigger is detected until the measurement is
performed.
Preset
0 s
Range
0 – 1 s
Default Unit
Seconds
Initial S/W
Revision
A.14.00
Trigger Level
Example
CALC:FPOW:POW1:DEF "TriggerLevel=2"
Notes
The trigger level parameter sets the voltage value at which an external trigger is detected.
Preset
1.2 V
Range
–5 to 5 V
Default Unit
Volts
Initial S/W Revision
A.14.00
Trigger Slope
Example
CALC:FPOW:POW1:DEF "TriggerSlope=Negative"
Notes
The trigger slope parameter indicates the direction of the edge trigger voltage for detection.
Preset
Positive
Range
Positive, Negative
Initial S/W Revision
A.14.00
2658
W-CDMA & HSPA Measurement Applications Reference
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...