4 Input/Output Functions
Input/Output
(s) the reference input types (in SCPI commands, and in the Virtual Front Panel menus ) are blanked and
unavailable for use.
Key Path
Input/Output
Remote Command
[:SENSe]:ROSCillator:SOURce:TYPE INTernal | EXTernal | SENSe | PULSe
[:SENSe]:ROSCillator:SOURce:TYPE?
Dependencies
The PULSe parameter, and support of the 1 pps signal at the EXT REF IN input, are not available in
firmware prior to A.13.00. They are also not available in some model numbers. If not available, the
Pulse key will be blank, and sending the PULSe parameter via SCPI will generate an error:
Preset
This is unaffected by a Preset but is set to SENSe on a "Restore Input/Output Defaults" or "Restore
System Defaults->All".
State Saved
Saved in instrument state.
Status Bits/OPC
dependencies
STATus:QUEStionable:FREQuency bit 1 set if unlocked.
Note: The status bit is not set for non-controlling instances. To determine if the frequency reference
is unlocked, the controlling instance must be queried.
Backwards Compatibility
Notes
Freq Ref In was not saved in state in the legacy instruments. It is a part of state in the X-Series.
Initial S/W Revision
Prior to A.02.00
Modified at S/W Revision
A.14.00
Remote Command
[:SENSe]:ROSCillator:SOURce?
Notes
The query [SENSe]:ROSCillator:SOURce? returns the current switch setting. This means:
1. If it was set to SENSe but there is no external reference nor 1pps signal so the instrument is
actually using the internal reference, then this query returns INTernal and not SENSe.
2. If it was set to SENSe and there is an external reference present, the query returns EXTernal and
not SENSe.
3. If it was set to SENSe and there is a 1 pps signal present, the query returns PULSe and not SENSe.
4. If it was set to EXTernal, then the query returns "EXTernal"
5. If it was set to INTernal, then the query returns “INTernal”.
6. If it was set to PULSe, then the query returns “PULSe”
Note: The SCPI query always returns “INTernal” for non-controlling instances.
Preset
SENSe
Backwards Compatibility
Notes
The query [:SENSe]:ROSCillator:SOURce? was a query-only command in ESA which always returned
whichever reference the instrument was using. The instrument automatically switched to the ext ref
if it was present.
In PSA (which had no sensing) the command [:SENSe]:ROSCillator:SOURce set the reference (INT or
EXT), so again its query returned the actual routing.
Thus the query form of this command is 100% backwards compatible with both instruments.
Initial S/W Revision
Prior to A.02.00
Remote Command
[:SENSe]:ROSCillator:SOURce INTernal|EXTernal
W-CDMA & HSPA Measurement Applications Reference
217
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...