12 Code Domain Measurement
Single (Single Measurement/Sweep)
Single (Single Measurement/Sweep)
Sets the analyzer for Single measurement operation. The single/continuous state is Meas Global, so the
setting will affect all the measurements. If you are Paused, pressing Single does a Resume.
See
"More Information" on page 1808
Key Path
Front-panel key
Example
:INIT:CONT OFF
Notes
See Cont key description.
Backwards Compatibility
Notes
For Spectrum Analysis mode in ESA and PSA, the Single hardkey and the INITiate:IMM switched
from continuous measurement to single measurement and restarted sweeps and averages (displayed
average count reset to 1), but did not restart Max Hold and Min Hold. In the X-Series, the Single
hardkey and the INITiate:IMM command initiate a sweep/ measurement/ average sequence/hold
sequence including MaxHold and MinHold.
For Spectrum Analysis mode in ESA and PSA, the Single hardkey restarted the sweep regardless of
whether or not you were in an active sweep or sweep sequence. In the X-Series, Restart does this
but Single only restarts the sweep or sweep sequence if you are in the idle state.
INIT[:IMM] in ESA & PSA Spectrum Analysis Mode does an implied ABORt. In some other PSA Modes,
INIT[:IMM] is ignored if not in the idle state. . The X-Series follows the ESA/PSA SA Mode model,
which may cause some Modes to have compatibility problems.
Initial S/W Revision
Prior to A.02.00
More Information
See
for details on the INIT:IMMediate (Restart) function.
If you are already in single sweep, the INIT:CONT OFF command has no effect.
If you are already in Single Sweep, then pressing the Single key in the middle of a sweep does not restart
the sweep or sequence. Similarly, pressing the Single key does not restart the sweep or sequence if the
sweep is not in the idle state (for example, if you are taking a very slow sweep, or the analyzer is waiting for
a trigger). Instead, it results in a message. "Already in Single, press Restart to initiate a new sweep or
sequence". Even though pressing the Single key in the middle of a sweep does not restart the sweep,
sending INIT:IMMediate does reset it.
To take one more sweep without resetting the average count, increment the average count by 1, by
pressing the step up key while Average/Hold Number is the active function, or sending the remote
command CALC:AVER:TCON UP.
1808
W-CDMA & HSPA Measurement Applications Reference
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...