12 Code Domain Measurement
Source
Source
Opens a menu of keys that access various source configuration menus and settings. In the test set,
pressing this key also causes the central view area to change and display the Source Control Main view.
Key Path
Front-panel key
RF Output
This parameter sets the source RF power output state.
Key Path
Source
Remote Command
:OUTPut[:EXTernal][:STATe] ON|OFF|1|0
:OUTPut[:EXTernal][:STATe]?
Example
OUTP OFF
OUTP?
Notes
The EXTernal node is shown in RD text so the SCPI remains the same between internal and external
source control. However, for EXT we do not wish to document this node to the customer since we
are controlling the internal source rather than the external source.
This setting is for the independent mode and has no effect on the
. If
the
is set to ON, the list sequencer controls the source output and this
key will be grayed-out. And this setting will be none-forceful grey out on front panel to indicate out-
of-scope. Non-forceful means user still can change this setting by SCPI but cannot change on front
panel. When set to OFF will make source leave list sequencer and this setting will be black out and
take effect immediately.
When the RF Output is ON, an “RF” annunciator is displayed in the system settings panel. When the
RF Output is turned Off, the RF annunciator is cleared. If the
is set to ON,
the “RF” annunciator will be replaced by “SEQ” in the system settings panel, indicating that the
output is controlled by the list sequencer.
Preset
Off
Range
On | Off
Initial S/W Revision
A.05.00
Amplitude
Allows you to access the Amplitude sub-menu.
Key Path
Source
Notes
The sub-menu under this button is for independent mode and has no effect on
is set to ON, the list sequencer controls the source
output and this key will be grayed-out on front panel to indicate out-of-scope. When you set
to Off will make source leave list sequencer and this button will be black
out.
Initial S/W Revision
A.05.00
W-CDMA & HSPA Measurement Applications Reference
1809
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...