14 QPSK EVM Measurement
Source
legal step number will be updated.
Dependencies
The range is 1 to 1000 which is determined by the number of steps you have configured. For details
see
"Number of Steps" on page 2803
Initial S/W Revision
A.09.40
Step Configuration of Power parameter list (Remote Command Only)
This SCPI command is to configure “Power”parameter array of the whole List Sequencer at one time. The
number of array is same as step number defined in
"Number of Steps" on page 2803
Number of Steps. As a
step is setup, the value entered run through several levels of validation.
Remote Command
:SOURce:LIST:SETup:AMPLitude <ampl>, <ampl>, <ampl>, …
:SOURce:LIST:SETup:AMPLitude?
Example
:SOUR:LIST:SET:AMPL –50dBm,–40dBm,–30dBm
:SOUR:LIST:SET:AMPL?
Notes
The command is to setup below parameter array of whole list sequence.
Power <
ampl
> – specifies the output power for the step in dBm. For details of the valid ranges see
If input parameter number exceeds the step number defined by
Number of Steps then generate error ", and only those parametes whose index number falls in
legal step number will be updated.
Dependencies
The range is 1 to 1000 which is determined by the number of steps you have configured. For details
see
"Number of Steps" on page 2803
Initial S/W Revision
A.09.40
Step Configuration of Waveform parameter list (Remote Command Only)
This SCPI command is to configure “Waveform” parameter array of the whole List Sequencer at one time.
The number of array is same as step number defined in
"Number of Steps" on page 2803
Number of Steps.
As a step is setup, the value entered run through several levels of validation.
Remote Command
:SOURce:LIST:SETup:WAVeform <string>, <string>, <string>, …
:SOURce:LIST:SETup:WAVeform?
Example
:SOUR:LIST:SET:WAV “CW”,”Off”,”CONT”
:SOUR:LIST:SET:WAV?
Notes
The command is to setup below parameter array of whole list sequence.
Waveform <
string
> – specifies the waveform for playback during the step. The step can output either
a new ARB waveform, continue playback of the previous waveform, or output a CW tone. The options
for specifying these are:
<filename> - plays the specified waveform from the start. The filename value is the name of the file
within ARB playback memory, it is does not include the windows path to the file on the HDD. If you
enter a filename for a waveform that does not reside within ARB playback memory, an error is
generated.
W-CDMA & HSPA Measurement Applications Reference
2585
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...