![Keysight E6640A Measurement Applications User & Programmer Reference Download Page 2656](http://html1.mh-extra.com/html/keysight/e6640a/e6640a_measurement-applications-user-and-programmer-reference_19616452656.webp)
15 Power Control Measurement
Meas
Default Unit
dB
Initial S/W
Revision
A.14.00
IF Gain
Example
CALC:FPOW:POW1:DEF "IFGain=10"
Notes
The IF gain parameter allows the user to specify the gain at the IF stage anywhere from –6 to 16 dB (1 dB
steps). This is an advanced feature, and for most cases this should remain at its default value of 0 dB.
Preset
0 dB
Range
–6 – 16 dB (1 dB steps)
Default Unit
dB
Initial S/W
Revision
A.14.00
IF Type
Example
CALC:FPOW:POW1:DEF "IFType=B25M"
Notes
The IF type parameter allows the user to select between different IF paths. For example, if the signal is less
than 25 MHz wide, then the user can select the B25M path to take advantage of additional filtering on this
analog IF path.
Preset
B40M
Range
B10M, B25M, B40M
Initial S/W
Revision
A.14.00
Include Power Spectrum
Example
CALC:FPOW:POW1:DEF "IncludePowerSpectrum=True"
Notes
The power spectrum parameter allows the user to read data on the entire spectrum for diagnostic purposes.
It is not recommended for production use. See CALC:FPOW:POW[n]:READ2? for details on the binary format
of the response.
Preset
False
Range
True (return both channel power and full power spectrum) or False (returns only channel power)
Default Unit
Boolean
Initial S/W
Revision
A.14.00
2656
W-CDMA & HSPA Measurement Applications Reference
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...