13 Modulation Accuracy (Composite EVM) Measurement
User Preset
•
Clears the input and output buffers.
•
Sets the Status Byte to 0.
Key Path
User Preset
Remote Command
:SYSTem:PRESet:USER
Example
:SYST:PRES:USER:SAVE:SYST:PRES:USER
Notes
:SYST:PRES:USER:SAVE is used to save the current state as the user preset state.
Clears all pending OPC bits. The Status Byte is set to 0.
Pressing the User Preset front-panel key while already in the User Preset menu will cause the User
Preset to get executed
Couplings
A user preset will cause the currently running measurement to be aborted and cause the saved
measurement to be active. Recalling a User Preset file has the same issues that recalling a Save
State file has. Some settings may need to be limited and therefore re-coupled, since the capabilities
of the mode may have changes when the User Preset file was last saved.
Initial S/W Revision
Prior to A.02.00
User Preset All Modes
Recalls all of the User Preset files for each mode, switches to the power-on mode, and activates the saved
measurement from the power-on mode User Preset file.
When the instrument is secured, all of the user preset files are converted back to their default user preset
files.
The User Preset function does the following:
•
Aborts the currently running measurement.
•
Switches the Mode to the power-on mode.
•
Restores the User Preset files for each mode.
•
Makes the saved measurement for the power-on mode the active measurement.
•
Brings up the saved menu for the power-on mode.
•
Clears the input and output buffers.
•
Sets the Status Byte to 0.
Key Path
User Preset
Remote Command
:SYSTem:PRESet:USER:ALL
Example
:SYST:PRES:USER:SAVE:SYST:PRES:USER:ALL
Notes
Clears all pending OPC bits. The Status Byte is set to 0.
:SYST:PRES:USER:SAVE is used to save the current state as the user preset state.
Couplings
A user preset will cause the currently running measurement to be aborted, cause a mode switch to
the power-on mode, and cause the saved measurement to be active in the power-on mode.
W-CDMA & HSPA Measurement Applications Reference
2311
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...