9 ACP Measurement
Source
:SOURce:RADio:ARB:RETRigger?
Example
:SOUR:RAD:ARB:RETR OFF
Notes
ON: Buffered Trigger
OFF: No Retrigger
Immediate: Restart on Trigger
This is defined as an enumerated SCPI command, with ON|OFF being considered as enumerated
types rather than Boolean. This means the query will return OFF instead of 0, and ON instead of 1.
Preset
ON
Range
No Retrigger | Buffered Trigger | Restart on Trigger
Initial S/W Revision
A.05.00
No Retrigger
Selects No Retrigger as the trigger response for single trigger type. No Retrigger sets the waveform
generator to play a waveform sequence or segment once when a trigger is received. Any triggers then
received during playback are ignored.
Key Path
Source, Modulation Setup, ARB, Trigger Type, Single
Example
:SOUR:RAD:ARB:RETR OFF
Initial S/W Revision
A.05.00
Buffered Trigger
Selects Buffered Trigger as the trigger response for single trigger type. Buffered Trigger sets the waveform
generator to play a waveform sequence or segment once when a trigger is received. If a trigger is received
during playback, the waveform generator plays the sequence or segment to the end, then plays the
sequence or segment once more.
Key Path
Source, Modulation Setup, ARB, Trigger Type, Single
Example
:SOUR:RAD:ARB:RETR ON
Initial S/W Revision
A.05.00
Restart on Trigger
Selects Restart on Trigger as the trigger response for single trigger type. Restart on Trigger sets the
waveform generator to play a waveform sequence or segment once when a trigger is received. If a trigger
is received during playback, the waveform generator resets and plays the sequence or segment from the
start.
Key Path
Source, Modulation Setup, ARB, Trigger Type, Single
Example
:SOUR:RAD:ARB:RETR IMM
Initial S/W Revision
A.05.00
W-CDMA & HSPA Measurement Applications Reference
845
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...