8 Channel Power Measurement
Source
Key Path
Source, Modulation Setup, ARB, ARB Setup
Remote Command
:SOURce:RADio:ARB:SCLock:RATE <freq>
:SOURce:RADio:ARB:SCLock:RATE?
Example
:SOUR:RAD:ARB:SCL:RATE 48.00 MHz
Notes
If there is a sample rate specified in the header of the waveform file, changing that sample rate is not
recommended, as it may cause problems with burst timing.
Dependencies
When a new waveform is selected for playback the settings contained within the associated
waveform header file are applied to the ARB. The sample rate is one of the values stored within the
header file. If the newly selected waveform file has an associated header file, the sample rate is
updated with the value from the header file. The sample rate will remain unchanged if the newly
selected waveform does not have an associated header file.
Preset
125.00 MHz
Min
1.00 kHz
Max
125.00 MHz
Initial S/W Revision
A.05.00
Run-Time Scaling
Allows you to adjust the run-time scaling value. The run-time scaling value is applied in real-time while the
waveform is playing.
Key Path
Source, Modulation Setup, ARB, ARB Setup
Remote Command
:SOURce:RADio:ARB:RSCaling <real>
:SOURce:RADio:ARB:RSCaling?
Example
:SOUR:RAD:ARB:RSC 100.00
Notes
This setting cannot be set in E6640A. Grey out on menu and the value is fixed at 70.00%.
Dependencies
When a new waveform is selected for playback the settings contained within the associated
waveform header file are applied to the ARB. The run-time scaling is one of the values stored within
the header file. If the newly selected waveform file has an associated header file, the run-time
scaling is updated with the value from the header file. The run-time scaling will remain unchanged if
the newly selected waveform does not have an associated header file.
Preset
70.00 %
Min
1.00 %
Max
100.00 %
Initial S/W Revision
A.05.00
Baseband Freq Offset
Allows you to adjust the value by which the baseband frequency is offset relative to the carrier.
W-CDMA & HSPA Measurement Applications Reference
509
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...