12 Code Domain Measurement
Meas Setup
1 frame |
2 frames |
3 frames |
4 frames (Long Mode) |
8 frames (Long Mode) |
16 frames (Long Mode)
Initial S/W Revision
Prior to A.02.00
Frame Capture Interval (Remote Command only)
Sets the data capture length in frames (1 frame equals 10 ms) that will be used in the acquisition.
Mode
WCDMA
Remote Command
[:SENSe]:CDPower:CAPTure:TIME[:FRAMe] <real>
[:SENSe]:CDPower:CAPTure:TIME[:FRAMe]?
Example
CDP:CAPT:TIME 1
CDP:CAPT:TIME?
Preset
1
State Saved
Saved in instrument state.
Range
0.067 (any value below 1 is rounded to the nearest multiple of 0.067), 1.0, 2.0, 3.0, 4.0 and 8.0 and
16.0 frames (0.67 to 80 ms; 1/15 frame equals 1 slot). Other numeric values between 1 and 16 are
rounded to the nearest integer; entries between integers are rounded up, excepting for entries above
16, which are rounded down to 16.
Min
0.067
Max
16.0
Initial S/W Revision
Prior to A.02.00
Symbol Analysis
Accesses the Symbol Analysis Menus.
Key Path
Meas Setup
Initial S/W Revision
Prior to A.02.00
tDPCH (BTS Only)
Toggles between Auto and Man. tDPCH value can be set when manual is selected. When AUTO is
selected, tDPCH value is shown as "---". tDPCH specifies chip offset (1 unit = 256 chips) of DPCH slot #0
start point from CPICH slot #0 start position. tDPCH value affects the measurement results such as
Symbol EVM (RMS, Peak), Symbol Magnitude Error, Symbol Phase Error, and Channel Power (Average
1748
W-CDMA & HSPA Measurement Applications Reference
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...