![Keysight E6640A Measurement Applications User & Programmer Reference Download Page 173](http://html1.mh-extra.com/html/keysight/e6640a/e6640a_measurement-applications-user-and-programmer-reference_1961645173.webp)
3 Programming the Test Set
STATus Subsystem
Preset
0
Min
0
Max
32767
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Integrity Uncalibrated Positive Transition
This command determines which bits in the Questionable Integrity Uncalibrated Condition register will set
the corresponding bit in the Questionable Integrity Uncalibrated Event register when the condition register
bit has a positive transition (0 to 1). The variable <integer> is the sum of the decimal values of the bits that
you want to enable.
Mode
All
Remote Command
:STATus:QUEStionable:INTegrity:UNCalibrated:PTRansition <integer>
:STATus:QUEStionable:INTegrity:UNCalibrated:PTRansition?
Example
STAT:QUES:INT:UNC:PTR 1 Oversweep (Meas Uncal) occurred will be reported to the Integrity
Summary of the Status Questionable register.
Preset
32767
Min
0
Max
32767
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Power Register
"Questionable Power Condition " on page 173
"Questionable Power Enable " on page 174
"Questionable Power Event Query " on page 174
"Questionable Power Negative Transition " on page 175
"Questionable Power Positive Transition " on page 175
Questionable Power Condition
This query returns the decimal value of the sum of the bits in the Questionable Power Condition register.
The data in this register is continuously updated and reflects the current conditions.
W-CDMA & HSPA Measurement Applications Reference
173
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...