Model 4200A-SCS Parameter Analyzer Reference Manual
Section 6: Clarius
4200A-901-01 Rev. C / February 2017
6-33
Parameter
Description
(on page 6-43)
Available when Pulse Mode is selected. The amount of time that the pulse is
on (5 ms to 20 s).
Available when Pulse Mode is selected. The amount of time that the pulse is
off (5 ms to 20 s).
(on page 6-43)
Available when Pulse Mode is selected. The current level that the
instrument sources when the pulse output is off. The level that you can set
depends on the present source range.
A limit that restricts the maximum voltage level that the instrument can
source. It is in effect when either current or voltage is sourced.
When this box is selected, the instrument measures current. Current is
recorded in the Analyze sheet and shown in the graph.
Report Value (Report Current or
Measure Current)
(on page 6-46)
The Report Value setting determines which current values are recorded in
the Analyze spreadsheet.
(on page 6-
The name of the current measurement. This is the name that Clarius
displays in the Analyze spreadsheet column for that measurement.
(on page 6-
When this box is selected, the instrument measures voltage. Voltage is
recorded in the Analyze sheet and shown in the graph.
(on page 6-
The name of the voltage measurement. This is the name that Clarius
displays in the Analyze spreadsheet column for that measurement.
The measure range determines the full-scale measurement span that is
applied to the signal.
(on page 6-46)
When this option is selected, Clarius records measurement status
information when the test executes. A column of the Analyze spreadsheet
displays this information. Hover over a cell to review the information.
Current Log Sweep operation mode - SMU
The Current Log Sweep operation mode allows you to sweep over a large range and plot the
measurements on a logarithmic scale.
An example of a log sweep is shown in the following figure.
Figure 230: Example log sweep