Model 4200A-SCS Parameter Analyzer Reference Manual
Section 9: Keithley External Control Interface (KXCI)
4200A-901-01 Rev. C / February 2017
9-35
ME
This command controls measurements.
Usage
ME
A
A
Measurement action:
•
SingleTrigger test, store readings in cleared buffer:
1
•
RepeatTrigger test, store readings in cleared buffer:
2
•
AppendTrigger test, append readings to buffer:
3
•
StopAbort test:
4
Details
The
ME1
or
ME2
command triggers the start of the test and makes the programmed number of
measurements. The measured readings are stored in the buffer. Note that the buffer is cleared before
readings are stored.
The
ME3
command also triggers the test but does not clear the buffer before storing the measured
readings. The readings are appended to the readings already stored in the buffer. The buffer can hold
up to 4096 readings.
The
ME4
command aborts the test.
Example
ME1
This command string triggers the start of the test and stores the readings in the cleared buffer.
Also see
None
Data output and file commands
The command strings for the following operations are valid in any system mode page:
•
Obtain output data
•
Save file
•
Get file