Appendix K: Using a Signatone CM500 Prober
Model 4200A-SCS Parameter Analyzer Reference Manual
K-10
4200A-901-01 Rev. C / February 2017
3. Move the mouse onto the WAFERMAP window and then do one of the following actions:
•
Select the dies to be tested on the wafermap and click
Enter
.
•
Click
Enter All
to test all dies.
Figure 812: CM500 Prober wafermap includes program sites
4. To step through all the programmed sites, select the
Run Program Site
icon on the toolbar.
Figure 813: CM500 Prober Run Program Sites icon
5. Click the
To First Site
button to move the prober to the first programmed site for testing. Make
sure the Subsite (template) is disabled here.
Figure 814: CM500 Prober Run Program Site window