Model 4200A-SCS Parameter Analyzer Reference Manual
Appendix B: Using a Model 590 C-V Analyzer
4200A-901-01 Rev. C / February 2017
B-7
Place capacitance source values in a spreadsheet (display-cap-file)
To place the source values in the Analyze sheet:
1. Select
display-cap-file
. The default parameters are displayed, as shown in the following figure.
Figure 553: DisplayCableCompCaps590 default parameters
2. Ensure that the
CabCompFile
field has the same file directory path that is used in
save-cap-
file
Enter and save capacitance source values (save-cap-file)
(on page B-6)).
3. Set the array size in the
RangeSize
,
Values100kSize
, and
Values1MSize
fields.
4. Click
Run
. The calibration source values are placed into the Analyze sheet for this action.
5. Select
Analyze
to view the spreadsheet for
display-cap-file
. An example spreadsheet is
shown in the following figure.
Figure 554: Display-cap-file spreadsheet showing capacitor source values