Section 8: Keithley User Library Tool (KULT)
Model 4200A-SCS Parameter Analyzer Reference Manual
8-14
4200A-901-01 Rev. C / February 2017
To copy libraries and user modules:
1. Copy the files from the 4200 to the 4200A-SCS.
2. Start KULT (
Start > Programs > Keithley Instruments > KULT
).
3. Select
File > Open Library
.
4. Select the library.
5. Select
File > Open Module
.
6. Select a module and select
OK
.
7. In the
Options
menu, click
Compile
.
8. Select the
Build
tab and verify that compilation was successful.
Figure 426: Successful compilation message displayed in Build tab
9. Open and compile any other user modules in this library.
10. From the
Options
menu, select
Build Library
. The user library is built. All of the user modules in
the presently open user library, and any libraries on which the presently open user module
depends, are linked together. A DLL is created that is accessible using UTMs in Clarius.
11. To test the new module, refer to
(on page 6-108).
Enabling real-time plotting for UTMs
To enable real-time plotting in a UTM, you use the following LPT library functions:
•
(on page 13-33)
•
(on page 13-34)
•
(on page 13-35)
The first parameter in the three functions is the variable name, defined as
char *
.
When using the new functions to transfer data into the data sheet in real time, make sure the data is
already in the memory of the 4200A-SCS. Sweep measurements are not suitable for real-time
transfer because data is not ready until sweep finishes. The following tutorials show how to enable
real-time plotting for a UTM.