Model 4200A-SCS Parameter Analyzer Reference Manual
Section 6: Clarius
4200A-901-01 Rev. C / February 2017
6-129
SegARBConfig control parameters
The following tables summarize all the parameters in the
SegARBConfig
control. In the tables, each
parameter is shown mapped to its parameter target. As mentioned above, the Segment Arb
functionality is primarily in two LPT commands:
seg_arb_sequence
and
seg_arb_waveform
. In
addition, a few parameters are unique to the
SegARBConfig
control.
SegARBConfig parameter names mapped to LPT function seg_arb_sequence
In this table, the variable name is in the function
(on page 13-125).
Parameter
name
Variable name in
function
Type
Required in
SegARBConfig?
Description
SegTime
Time
Double array
Yes
Time duration for each
segment
SegStartV
StartV
Double array
Yes
Segment start voltage
SegStopV
StopV
Double array
Yes
Segment stop voltage
SegSSR
SSR
Integer array
Yes
Solid State Relay control, per
segment
SegTrig
Trig
Integer array
Yes
Trigger output state for each
segment
SegMeasType
MeasType
Integer array or
Integer
1
No
Measurement Type: None,
spot mean, or waveform
SegMeasStart
MeasStart
Double array or
Double
1
No
Start point for measurement
window, for each segment;
from 0 to 1 (100%)
SegMeasStop
MeasStop
Double array or
Double
1
No
Stop point for measurement
window, per segment; from 0
to 1 (100%)
NumSegments
NumSegments
Integer
No
Number of segments in a
sequence
1
If using a single value instead of an array, the
SegARBConfig
control automatically assigns the value to each
segment.
SegARBConfig parameter names mapped to SegARBConfig
In this table, the variable name is in the function
SegARBConfig
.
Parameter
name
Variable name in
function
Type
Required in
SegARBConfig?
Description
SegMeas
SegMeas
Integer array
No
Measurement Enabled
1
SeqList
SeqList
Integer array
No
List of defined sequences in
SegARBConfig
SeqStartSeg
SeqStartSeg
Integer array
No
Start segment array value in
SegARBConfig
SeqStopSeg
SeqStopSeg
Integer array
No
Stop segment array value in
SegARBConfig
1
SegMeas
parameter turns a measurement on or off for each segment and is independent of the
SegMeasType
(based on implementation in user module).