Model 4200A-SCS Parameter Analyzer Reference Manual
Section 13: LPT library function reference
4200A-901-01 Rev. C / February 2017
13-127
This figure shows an example of a Segment Arb sequence defined by the
seg_arb_sequence
command. Spot mean discrete measurements are performed on segments two and four.
Figure 508: Segment Arb sequence example
This table lists the
seg_arb_sequence
parameter arrays for the Segment Arb sequence shown in
the example.
Parameter arrays for the
seg_arb_sequence
example
Parameter
Array Name
Value
SegNum
Seg_Num
1
2
3
4
5
6
7
StartV
Start_Volt
0 V
1 V
1 V
1.5 V
1.5 V
0 V
0 V
StopV
Stop_Volt
1 V
1 V
1.5 V
1.5 V
0 V
0 V
0 V
Time
Time_Interval
50e-9 s
100e-9 s
20e-9 s
150e-9 s
50e-9 s
500e-9 s
130e-9 s
Trig
Trigger_Level
1 (high)
1 (high)
1 (high)
0 (low)
0 (low)
0 (low)
0 (low)
SSR
Output_Relay
1 (closed)
1 (closed)
1 (closed)
1 (closed)
1 (closed)
1 (closed)
0 (open)
MeasType
Meas_Type
0 (none)
1 (spot mean) 0 (none)
1 (spot mean)
0 (none)
0 (none)
0 (none)
MeasStart
Meas_Start
0 s
25e-9 s
0 s
50e-9 s
0 s
0 s
0 s
MeasStop
Meas_Stop
0 s
75e-9 s
0 s
100e-9 s
0 s
0 s
0 s
Also see
(on page 5-56)
(on page 13-128)