Appendix B: Using a Model 590 C-V Analyzer
Model 4200A-SCS Parameter Analyzer Reference Manual
B-28
4200A-901-01 Rev. C / February 2017
Csize
Set to a value equal to or greater than the number of voltage steps in the sweep or
equal to ((
LastBias
−
FirstBias
) /
StepV
) + 1
V
Output: The array of bias voltages used
Vsize
Set to a value equal to or greater than the number of voltage steps in the sweep or
equal to ((
LastBias
−
FirstBias
) /
StepV
) + 1
G_or_R
Output:
•
When the parallel measurement model (0) is selected,
G_or_R
is the
measured conductance
•
When the series measurement model (1) is selected, this is the measured
resistance
G_or_Rsize
When this function is called from a Clarius UTM, this parameter is fixed at 1350
T
The array of time stamps for each measurement step
Tsize
When this function is called from a Clarius UTM, this parameter is fixed at 1350
Details
This user module performs a capacitance versus voltage staircase sweep. The figure below shows
the default parameters for the
590-cvsweep
UTM, which uses the
CvSweep590
user module.
Figure 566: CvSweep590 user module (590-cvsweep UTM)