Pulsers
4220-PGU
4225-PMU
Pulse modes
Source, Segment Arb
Details
You can configure each channel to output its own unique Segment Arb waveform. A Segment Arb
waveform is made up of user-defined segments. Each segment can have a unique time interval, start
value, stop value, output trigger level (TTL high or low), and output relay state (open or closed).
See
(on page 5-56) for details on this pulse mode. For triggering, refer to
(on page 5-67).
To configure each channel to output a unique Segment Arb
®
waveform, refer to
(on page 13-125).
The following arrays are required for the example Segment Arb waveform shown here.
Figure 514: Segment Arb sequence example