Model 4200A-SCS Parameter Analyzer Reference Manual
Appendix D: Using a Model 82 C-V System
4200A-901-01 Rev. C / February 2017
D-11
2. Ensure that the CabCompFile field has the same file directory path that is used in
capacitance source values (SaveCableCompCaps82)
(on page D-10).
3. Set the other fields to
8
.
4. Click
Run
. The calibration source values entered into the action are placed into its spreadsheet.
5. Select
Analyze
. The sheet displays the values. An example spreadsheet is shown here.
Figure 587: display-cap-file spreadsheet with capacitor source values
Compensate for cable capacitance (CableCompensate82)
To compensate for cable capacitance:
1. In the project tree, select
cable-compensate
or
cablecomp
.
2. Select
Configure
.
3. Ensure that the
CabCompFile
field of the parameter list has the same file directory path that is
used in
Enter and save capacitance source values
(on page D-10).
4. Enable or disable cable compensation: Use the frequency and range fields to either disable (
0
) or
enable (
1
) cable compensation for the test frequencies and ranges. The following figure shows
cable compensation enabled for all ranges and test frequencies.
Figure 588: CableCompensate82 user module