Section 6: Clarius
Model 4200A-SCS Parameter Analyzer Reference Manual
6-160
4200A-901-01 Rev. C / February 2017
Set up Linear Stress/Measure timing
To set timing for linear stress/measure mode:
1. Select
Linear
.
2. In
First Stress Time
, set the time in seconds that devices are stressed during the first cycle.
3. In
Last Stress Time
, set the total stress time for the subsite. This is the total amount of stress
time that will have passed when the last stress is complete.
4. In
Number of Stresses
, set the total number of stresses. You can set up to 128 stresses.
5. If needed, enter the
Stress/Measure Delay
in seconds. This is the delay after each stress cycle.
It allows the device to reach equilibrium before the next measurement.
Clarius uses the values you entered to calculate the stress times for the cycles. The times are
displayed in the Stress Times box in seconds.
Figure 323: Set timing for linear stress/measure mode