9-51
System Mode SMU default settings
When KXCI mode is started, the SMUs are in a default state that mimics the Keysight Model 4145B,
as shown in the channel definition and source setup tables below. This means that each SMU is
active and part of the test, whether or not the test used all the SMUs in the 4200A-SCS chassis.
This may be undesirable, as many tests use a different number of SMUs or SMUs in a different state
from the Keysight 4145 default. To avoid the complication of building a KXCI test starting with the
default setup, use the following commands to remove the SMUs from the setup:
DE CH1 CH2 CH3 CH4 VM1 VM2 VS1 VS2
The above shows the method for a system with eight SMUs. A system with only four SMUs would not
require the last four definitions:
DE CH1 CH2 CH3 CH4
Channel Definition (Page
DE
) KXCI defaults
Instrument
Name
Source
Command
V
I
Mode
Function
SMU1
SMU2
SMU3*
SMU4*
SMU5*
SMU6*
SMU7*
SMU8*
V1
V2
V3
V4
VS1
VS2
VM1
VM2
I1
I2
I3
I4
N/A
N/A
N/A
N/A
COM
I
V
V
V
V
N/A
N/A
CONSTant
VAR2
VAR1
CONSTant
CONSTant
CONSTant
N/A
N/A
CH
CH
CH
CH
VS
VS
VM
VM
* If there are less than eight SMUs, only the SMUs in the chassis
are programmed to the defaults shown.
Source Setup (Page
SS
) KXCI defaults
VAR1
(Command VR)
VAR2
(Command IP)
Name
V3
I2
Sweep mode
Linear
Linear
Start
0.00 V
20.00 µA
Stop
1.00 V
N/A
Step
0.010 V
20.00 µA
Number of steps
101
5
Compliance
100.0 mA
2 V
Constant
Source
Compliance
Command
V1 COM
V4 V
VS1 V
VS2 V
0.00 V
0.00 V
0.00 V
0.00 V
105.0 mA
100.0 mA
N/A
N/A
N/A
VC
SC
SC