1
Instruction Manual
HI
41
10
FLUORIDE COMBINA
TION
FLUORIDE
HI 4010
HI 4110
Fluoride Ion
Selective Electrode
Half-cell
Combination
Page 1: ...Instruction Manual HI 4110 FLUORIDE COMBINATION FLUORIDE HI 4010 HI 4010 HI 4010 HI 4010 HI 4010 HI 4010 HI 4110 HI 4110 HI 4110 HI 4110 HI 4110 Fluoride Ion Selective Electrode Half cell Combination...
Page 2: ...o forms HF species below pH 5 Increase pH in these cases above 5 for a total fluoride measurement Operating Temperature 0 80 C Operating pH 5 to 8 pH Dimensions 12 mm OD X 120 mm nominal insertion 0 4...
Page 3: ...standard and ISA from the list below Part number 0 1 M Sodium Fluoride 500 mL HI4010 01 100 ppm 500 mL HI4010 02 1000 ppm 500 mL HI4010 03 10 ppm with TISAB II 500 mL HI4010 10 1 ppm with TISAB II 500...
Page 4: ...ntact Hanna Instruments for guidelines with other applications VII VII VII VII VII General Guidelines General Guidelines General Guidelines General Guidelines General Guidelines Calibration standards...
Page 5: ...r is inverted an air pocket will be seen which is normal 3 Prepare HI 5315 reference electrode by filling outer electrolytereservoirwithHI7072 4 Place sensor and reference electrodes into electrode ho...
Page 6: ...uoridestandardfor1hour Repeatsection IX If the reading is jumpy or unstable shake sensor down seesectionVIII If the membrane is damaged the response becomes extremely sluggish or the slope of the elec...
Page 7: ...ution to condition before calibration XI XI XI XI XI Direct Calibration and Measurement Direct Calibration and Measurement Direct Calibration and Measurement Direct Calibration and Measurement Direct...
Page 8: ...and Care of the HI 4010 and Storage and Care of the HI 4010 and Storage and Care of the HI 4010 and Storage and Care of the HI 4010 and HI 4110 sensors HI 4110 sensors HI 4110 sensors HI 4110 sensors...
Page 9: ...r 6 monthsfromdateofpurchasewhenusedfortheirintended purpose and maintained according to instructions If they failtoworkwhenfirstusedcontactyourdealerimmediately Damageduetoaccidents misuse misapplica...