Model 4200A-SCS Parameter Analyzer Reference Manual
Appendix A: Using switch matrices
4200A-901-01 Rev. C / February 2017
A-27
5. From the
Model
list, select the prober.
6. Enter the
Number of Pins / Positioners
.
7. Select the options that are appropriate for your prober.
The number of pins defined in the probe station properties determines the pins that are available to
assign to a switch matrix card column. Make sure the number of pins assigned is appropriate for
your system.
Step 3. Add switching system mainframe
To add a switching system mainframe:
1. Select
Add External Instrument
.
2. Select the
Keithley 707/707A/707B Switching Matrix
or
Keithley 708/708A/708B Switching
Matrix
.
3. Select
OK
.
4. In the Configuration Navigator, select the switching matrix. The properties are displayed. The
following figure shows the properties for the 707/707A/707B. If the 708/708A/708B mainframe is
selected, there is only one switch card slot.
Figure 542: KCon MTRX1 Properties tab