Model 4200A-SCS Parameter Analyzer Reference Manual
Section 9: Keithley External Control Interface (KXCI)
4200A-901-01 Rev. C / February 2017
9-77
:CVU:CABLE:COMP:SHORT
This command performs short compensation and collects the short compensation cable data for the CVU.
Usage
:CVU:CABLE:COMP:SHORT
Also see
(on page 4-10)
:CVU:CABLE:COMP:LOAD
This command performs load compensation and collects the load compensation cable data for the CVU.
Usage
:CVU:CABLE:COMP:LOAD
Also see
(on page 4-10)
:CVU:CABLE:COMP:MEASCUSTOM
This command performs custom cable length compensation and collects the compensation cable data for the
CVU.
Usage
:CVU:CABLE:COMP:MEASCUSTOM
Also see
(on page 4-10)
:CVU:DCV:OFFSET
This command applies an offset value to the DC low terminal.
Usage
:CVU:DCV:OFFSET
offsetv
offsetv
-30 V to +30 V
Details
Voltage offset to apply while sampling Z-measurements.
Also see
None