Section 4: Multi-frequency capacitance-voltage unit
Model 4200A-SCS Parameter Analyzer Reference Manual
4-94
4200A-901-01 Rev. C / February 2017
solarcell connections
The next figures show the basic test configurations for this project:
•
I-V testing
: The basic configuration shows four-wire sensing using a 4200-SMU or 4210-SMU.
Four-wire sensing is needed to achieve optimum performance. See
(on page 3-1) for details on 4200-SMU connections.
•
C-V Testing
(on page 4-5) for details on connections. Use only the
supplied (red) 100 Ω SMA cables for connections to the 4210-CVU. Be sure that all used SMA
cables are the same length.
After making or changing connections, be sure to use the Confidence Check diagnostic tool and do
connection compensation tests. Refer to
(on page 4-10) for details.
Figure 140: Basic configurations to test PV cells for I-V testing