Model 4200A-SCS Parameter Analyzer Reference Manual
Section 4: Multi-frequency capacitance-voltage unit
4200A-901-01 Rev. C / February 2017
4-19
Figure 97: Real-Time Measurement dialog box
3. Select the parameters for which you want to return results.
4. Set the Speed, AC Drive Conditions, and DC Bias Conditions for the conditions you want to test.
5. Select
Run
. The results for the selected parameters are displayed at the top of the dialog box.
Confidence Check
Confidence Check is a diagnostic tool that allows you to check the integrity of open and short
connections and connections to a device under test (DUT). When the 4210-CVU is connected to the
DUT, the Confidence Check displays the measured readings in real time in the Messages area of
Clarius.
An open or short confidence check makes a measurement on the high and the low sides of the test
circuit.