devint
This command resets all active instruments in the system to their default states.
Usage
int devint(void);
Details
Resets all active instruments in the system to their default states. It clears the system by opening all
relays and disconnecting the pathways. Meters and sources are reset to their default states. Refer to
the hardware manuals for the instruments in your system for listings of available ranges and the
default conditions and ranges.
The
devint
command is implicitly called by the
execut
command.
To abort a running
pulse_exec
pulse test, see
dev_abort
.
devint
does the following:
1.
Clears all sources by calling
devclr
.
2.
Clears the matrix crosspoints by calling
clrcon
.
3.
Clears the trigger tables by calling
clrtrg
.
4.
Clears the sweep tables by calling
clrscn
.
5.
Resets GPIB instruments by sending the string defined with
kibdefint
.
6.
Resets the active instrument cards.
Instrument cards are reset in the following order:
1.
SMU instrument cards
2.
CVU instrument cards
3.
Pulse instrument cards (4225-PMU or 4220-PGU)
devint
is implicitly called by
execut
and
tstdsl
.
devclr
is implicitly called by
clrcon
.
The SMUs return to the following states:
•
100 µA and 10 V ranges
•
Autorange on
•
Voltage source
•
0 V DCV bias
The 4210-CVU returns to the following states:
•
30 mV
RMS
AC signal
•
0 V DCV bias
•
100 kHz frequency
•
Autorange on
•
Cable length compensation set to 0 m
•
Open/Short/Load compensation disabled