Model 4200A-SCS Parameter Analyzer Reference Manual
Section 6: Clarius
4200A-901-01 Rev. C / February 2017
6-181
4. Set
Current Site
to the site that you want to run.
5. Select
Run
. The Run icon changes as shown below. The active site, actions, and tests are listed
to the left of Run. The Stop icon changes to red.
Figure 344: Run icon while a test is running
When the site completes, a beep sounds and the run arrows around the Run icon are no longer
displayed.
Cycle a subsite
Subsite cycling allows you to repeatedly cycle through the subsite tests. The data for every repeated
test is acquired and placed in its Analyze Stress tab.
Measured readings (output values) can be exported from individual tests into the subsite.
To run cycling for a single subsite:
1. Set up the subsite as described in
(on page 6-148).
2. In the project tree, select the subsite.
3. Select
Run
.
To run cycling for multiple subsites:
1. Set up the subsite as described in
(on page 6-148).
2. In the project tree, select the project.
3. Make sure the subsites you want to run are checked in the project tree.
4. Select
Run
.