Model 4200A-SCS Parameter Analyzer Reference Manual
Appendix F: Set up a probe station
4200A-901-01 Rev. C / February 2017
F-13
PrSSMovNxt
In learn mode, the
PrSSMovNxt
command causes the prober to move to the next subsite. If needed,
you can specify the inker to fire before the move.
Usage
status
= PrSSMovNxt(int
ink_number
);
status
Returned values; refer to
Details
ink_number
The inkers to fire:
•
0
: No inker; move only
•
1
: 1
•
2
: 2
•
3
: 1, 2
•
4
: 3
•
5
: 1, 3
•
6
: 2, 3
•
7
: 1, 2, 3
•
8
: 4
•
9
: 1, 4
•
10
: 2, 4
•
11
: 1, 2, 4
•
12
: 3, 4
•
13
: 1, 3, 4
•
14
: 2, 3, 4
•
15
: All 4
Library
Dependencies: PRBCOM
Details
The
PrMovNxt
function returns the values:
•
1
: Success (
PR_OK
)
•
-1014
: Move fail
•
-1013
: Unintelligible response
•
-1011
: Bad mode
-1015
: Unexpected error
Example
status = PrSSMovNxt(0);
Also see
None