Model 4200A-SCS Parameter Analyzer Reference Manual
Section 6: Clarius
4200A-901-01 Rev. C / February 2017
6-11
Add a device and test to the project
You can add additional items to a project. Once a project is added from the library to the project tree,
it is copied from the project in the library, so you can make changes without affecting the original
project. The project in the project tree is is stored in My Projects.
This example shows you how to add a predefined test to the project. Predefined tests are configured
with commonly used parameter settings and a set of typical data. Once they are in a project, you can
change the parameters as needed. They can be an efficient way for you to add a test to your project.
You can use the basic procedure described here to find any items in the library.
To add a four-terminal MOSFET device and test to the project:
1. Change the library to
Tests
.
2. In the Filters pane, select
Transistor
and
4
Terminals.
3. In the search box, type
MOSFET
and select
Search
.
4. Scroll to the
MOSFET Drain Family of Curves (vds-id)
test.
5. Select
Add
. The selected test and the device is added to the project tree under the previous item
that was highlighted.
6. To move the device and test, drag the device to a new location.
7. Select
Save
.
Figure 218: Add a MOSFET test and device to the project
If the device for a test is not in the project tree, Clarius adds the appropriate device when you add a
test to the project tree. You can also add the device and test separately.