Model 4200A-SCS Parameter Analyzer Reference Manual
Section 8: Keithley User Library Tool (KULT)
4200A-901-01 Rev. C / February 2017
8-33
11. Enter the name
v_sweep_chk
for the UTM. You will use the
v_sweep_chk
UTM to execute the
VSweep user module.
12. Select
Configure
.
13. In the right pane, from the User Libraries list, select the
my_2nd_lib
library.
14. From the User Modules list, select the
v_sweep_chk
user module. A default schematic and
group of parameters are displayed for the UTM.
15. For Vforce, select
Enter Values
and enter the sweep values.
16. Select
Run
.
17. Select
Analyze
.
At the conclusion of execution, review the results in the Analyze worksheet. If you connected a 1
kΩ
resistor between SMU1 and GNDU, used the default UTM parameter values, and executed the UTM
successfully, the results should be similar to the results in the following figure.The current/voltage
ratio for each row of results should be approximately 1 mA / V.
Figure 442: Checking the VSweep user module