Model 4200A-SCS Parameter Analyzer Reference Manual
Section 9: Keithley External Control Interface (KXCI)
4200A-901-01 Rev. C / February 2017
9-83
GN
This command queries input or output parameter values, or both, by name for the last user module run in KXCI.
Usage
GN
ParameterName
GN
ParameterName
NumValues
ParameterName
The name of the parameter in the KULT module
NumValues
The number of values in an output array to be returned; see
Details
Details
NumValues
is only used for an output parameter that is an array. If
NumValues
is not used, one
value is returned.
Arrays are returned as a list of values separated by semicolons.
The value returned for an input parameter is the given value. The values returned for an output
parameter is the outcome of the test (for example, measured readings).
You can also query a parameter by specifying the corresponding number for the parameter in the
KULT module (see
(on page 9-84)).
Example
GN Vforce 11
This command queries all 11 test voltages (
Vforce
parameter) for the function that was run in the
page 9-82) example.
The following array of test voltages are returned and displayed in the KXCI console:
0; 0.5; 1.0; 1.5; 2.0; 2.5; 3.0; 3.5; 4.0; 4.5; 5.0
Also see
None