Accuracy

 (linearity, hysteresis, repeatability)

±0.25% of full scale ±1 least significant digit

DPG1001B

: ±0.10% of full scale ±1 least significant digit 

Sensor hysteresis:  ±0.015% FS, included in accuracy 
Sensor repeatability:  ±0.01% FS, included in accuracy

Display

 (update rate, type, size)

3 readings per second nominal display update 

Ranges up to 1999

: 3-1/2 digit LCD, 1/2” digit height 

Ranges 2000 and up

: 4 digit LCD, 0.4” H digits, alphanumeric 

display of units

BBL

: Red LED display backlighting

Controls & Functions

Ranges up to 1999

: Front pushbutton turns gauge on/off and 

starts shutoff timer. Turns on backlighting for 1 minute on BBL 
models.

Ranges 2000 and up

: Front pushbutton turns gauge on/off, 

starts shutoff timer. Pressing and holding front button zeros 
gauge. The zero feature is not used on absolute reference 
gauges.

Calibration

Ranges up to 1999

: Front potentiometers, non-interactive 

zero & span, ±10% range

Ranges 2000 and up

: Internal pushbuttons, non-interactive 

zero, midpoint, span, ±10% range. Calibration information 
stored in non-volatile memory.

Shutoff

Last character of model number on rear label indicates shutoff 
time. Contact Omega for other optional shutoff times.

-5

5 minute auto shutoff (standard) 

-NS

:  No auto shutoff, on/off via front pushbutton

Batteries

2 AA alkaline, included

Battery Life

1000B, 1001B, 1100B ranges to 1999 

Approx. 2500 hours

1000B, 1001B, 1100B ranges 2000 and up 

Approx. 2000 hours

1000BBL, 1100BBL ranges to 1999 

Approx. 180 hrs (backlight on when gauge is on)

1000BBL, 1100BBL 2000 and up 

Approx. 150 to 1500 hrs depending on backlight usage

Low Battery Indication

Low battery symbol on display when batteries must be 
replaced

Connection and Material

1/4” NPT male fitting, 316L stainless steel
All wetted parts are 316L stainless steel

Dimensions

 (see drawing on next page)

DPG1000, DPG1001 

3.38” W x 2.88” H x 1.65” D housing 
Add approx. 0.75” to height for pressure fitting

DPG1100 

3.5” W x 3.0” H x 2.0” D NEMA 4X housing 
Add approx. 0.75” to height for pressure fitting

Weight

 (approximate)

Gauge: 

9 ounces 

Shipping weight: 

1 pound

Housing Material & Color

DPG1000, DPG1001 

Extruded aluminum case, light gray epoxy powder coated, 
black ABS/polycarbonate bezel, front and rear gaskets, poly-
carbonate label

DPG1100 

Light gray ABS/polycarbonate NEMA 4X case, rear gasket, 
polycarbonate label

Overpressure, Burst, Vacuum

Ranges using 3000 psig sensor: 

5000 psig 

Ranges using 5000 psig sensor: 

7500 psig 

All others: 

2 X pressure range

3000 psi, 5000 psi, and 4 digit ranges 112.5% full scale out-
of-range display: 1– – – or I –.–.–.–
4 X sensor burst pressure rating, or 10,000 psi, whichever is 
less
Vacuum service: ±15 psig, 15 psig, 15 psia, 30 psia, 100 psig, 
100 psia, 200 psig sensors.  
3000 psi, 5000 psi, and 4 digit ranges: Under-range display 
(non-vacuum sensors): –Err

Environmental

Storage temperature: 

–40 to 203°F (–40 to 95°C)

Operating (3.5 digit versions): 

–40 to 185°F (–40 to 85°C)

Operating (4 digit versions): 

–4 to 185°F (–20 to 85°C)

Compensated range: 

32 to 158°F (0 to 70°C)

DPG1100B, 

DPG1100BBL

DPG1000B, 

DPG1001B, 

DPG1000BBL

DPG1000B, 

DPG1000BBL 

Ranges above 

2000

DPG1100B, 

DPG1100BBL 

Ranges abovve 

2000

PSI

Res

inHg/PSI

Res

mmH2O

Res

 3PSIG

.01  –30V15PSIG

.1  2000MMH2OG

1

 5PSIG

.01

 –30V100PSIG

.1  3500MMH2OG*

1

 15PSIA

.01  –30V200PSIG

.1

cmH2O

Res

 15PSIVAC

.01

inH2O

Res

 200CMH2OG

.1

 

±15PSIG

.1  85INH2OG

.1  350CMH2OG

1

 15PSIG

.01  140INH2OG

.1  1000CMH2OA

1

 30PSIA

.1  400INH2OA

1  1000CMH2OVAC

1

 30PSIG

.1  400INH2OVAC

1  ±1000CMH2OG

1

 60PSIG

.1  ±400INH2OG

1  1000CMH2OG

1

 100PSIA

.1  400INH2OG

1  2000CMH2OA

1

 –15V100PSIG

.1  850INH2OA

1  2000CMH2OG

1

 100PSIG

.1  850INH2OG

1

kPa

Res

 –15V200PSIG

.1

ftH2O

Res

 20KPAG

.01

 200PSIG

.1  7FTH2O

.01  35KPAG

.1

 300PSIG

1  12FTH2O

.01  100KPAA

.1

 500PSIG

1  35FTH2O

.1  100KPAVAC

.1

 1000PSIG

1  70FTH2O

.1  ±100KPAG

.1

 2000PSIG

1  140FTH2O

.1  100KPAG

.1

 3000PSIG

1  230FTH2O

1  200KPAA

.1

 5000PSIG

1  480FTH2O

1  200KPAG

.1

oz/in2

Res

 700FTH2O

1  400KPAG

1

 50ZING

.1  1150FTH2O

1  700KPAA

1

 80ZING

.1  2300FTH2O*

1  700KPAG

1

 240ZINA

1  4600FTH2O*

1 –100V700KPAG

1

 240ZINVAC

1  6900FTH2O*

1  1400KPAG

1

 ±240ZING

1

mmHg

Res

–100V1400KPAG

1

 240ZING

1  150MMHGG

.1  2000KPAG

1

 480ZINA

1  260MMHGG

1  3500KPAG*

1

 480ZING

1  760MMHGA

1  7000KPAG*

1

inHg

Res

 760MMHGVAC

1

MPa

Res

 6INHGG

.01  ±760MMHGG

1  1.4MPAG

.001

 10INHGG

.01  760MMHGG

1 –0.1V1.4MPAG

.001

 30INHGA

.1  1600MMHGA

1  2MPAG

.001

 30INHGVAC

.1  1600MMHGG

1  3.5MPAG

.01

 ±30INHGG

.1

Torr

Res

 7MPAG

.01

 30INHGG

.1  760TORRA

1  14MPAG

.01

 60INHGA

.1  760TORRVAC

1  20MPAG

.01

 60INHGG

.1  1600TORRA

1  35MPAG

.1

 120INHGG

.1

mbar

Res

g/cm2

Res

 200INHGA

.1  200MBARG

.1  200GCMG

.1

–30V200INHGG

.1  350MBARG

1  350GCMG

1

 200INHGG

.1  1000MBARA

1  1000GCMA

1

–30V400INHGG

1  1000MBARVAC

1  1000GCMVAC

1

 400INHGG

1  ±1000MBARG

1  ±1000GCMG

1

 600INHGG

1  1000MBARG

1  1000GCMG 

1

 1000INHGG

1  2000MBARA

1  2000GCMA

1

 2000INHGG

1  2000MBARG

1  2000GCMG

1

 4000INHGG*

1

atm

Res

bar

Res

kg/cm2

Res

 1ATMA

.001  1BARA

.001  1KGCMA

.001

 1ATMVAC

.001  1BARVAC

.001  1KGCMVAC

.001

 ±1ATMG

.001  ±1BARG

.001  ±1KGCMG

.001

 1ATMG

.001  1BARG

.001  1KGCMG

.001

 2ATMA

.001  2BARA

.001  2KGCMA

.001

 2ATMG

.001  2BARG

.001  2KGCMG

.001

 4ATMG

.01  4BARG

.01  4KGCMG

.01

 7ATMA

.01  7BARA

.01  7KGCMA

.01

 7ATMG

.01  7BARG

.01  7KGCMG

.01

 –1V7ATMG

.01  –1V7BARG

.01  –1V7KGCMG

.01

 14ATMG

.01  14BARG

.01  14KGCMG

.01

 –1V14ATMG

.01  –1V14BARG

.01  –1V14KGCMG

.01

 20ATMG

.01  20BARG

.01  20KGCMG

.01

 34ATMG

.1  35BARG

.1  35KGCMG

.1

 70ATMG

.1  70BARG

.1  70KGCMG

.1

 140ATMG

.1  140BARG

.1  140KGCMG

.1

 200ATMG

.1  200BARG

.1  200KGCMG

.1

 340ATMG

1  350BARG

1  350KGCMG

1

Ranges and Resolution 

Resolution is fixed as indicated in table. 
Resolution is limited by number of display digits. 2, 20, 200, and 2000 ranges display 1.999, 19.99, 199.9, and 1999.

Bold

 

Standard ranges. Contact Omega Engineering to order other ranges listed below.

G

 

Gauge reference pressure

VAC

 

Vacuum gauge, minus sign not used unless specified

A

 

Absolute reference (normally reads atmospheric pressure, reads zero at full vacuum)

 

DPG1001B not available in 3 psi, 5 psi, vacuum, bipolar ranges, and ranges with insufficient digits.

*

 

4 digit range. All others 3-1/2 digit.

Digital Pressure Gauges

DPG1000B, DPG1001B, DPG1000BBL, 

DPG1100B, DPG1100BBL

INSTRUCTION

SHEET

DPG1000B 

Battery Powered 

DPG1001B 

Battery Powered, High Accuracy 

DPG1000BBL 

Battery Powered w. Backlighting 

DPG1100B 

Battery Powered, NEMA 4X 

DPG1100BBL 

Battery Powered w. Backlighting, NEMA 4X

M3359/0818

TM

Shop online at omega.com  e-mail: [email protected] 

For latest product manuals: www.omegamanual.info

Pb

Lead 

Free

Summary of Contents for DPG1000B

Page 1: ... 2000 DPG1100B DPG1100BBL Ranges abovve 2000 PSI Res inHg PSI Res mmH2O Res 3PSIG 01 30V15PSIG 1 2000MMH2OG 1 5PSIG 01 30V100PSIG 1 3500MMH2OG 1 15PSIA 01 30V200PSIG 1 cmH2O Res 15PSIVAC 01 inH2O Res 200CMH2OG 1 15PSIG 1 85INH2OG 1 350CMH2OG 1 15PSIG 01 140INH2OG 1 1000CMH2OA 1 30PSIA 1 400INH2OA 1 1000CMH2OVAC 1 30PSIG 1 400INH2OVAC 1 1000CMH2OG 1 60PSIG 1 400INH2OG 1 1000CMH2OG 1 100PSIA 1 400IN...

Page 2: ... The actual pressure is displayed Attempting to zero the gauge with pressure greater than approximately 3 of full scale applied will result in an error condition and the display will alternately indicate E r r 0 and the actual measured pressure The gauge must be powered down to reset the error condition Absolute reference gauges do not use the zero feature since they read atmospheric pressure unde...

Page 3: ...s in the calibration mode until restarted manually or power is removed Features not related to calibration are disabled If the power is removed during calibration settings will not be saved The display will indicate the current pressure reading updating approximately 3 times per second Each press of the UP or DOWN button makes a small correction which may not always be indicated on the display Pre...

Page 4: ...by OMEGA if the unit is found to be defective it will be repaired or replaced at no charge OMEGA s WARRANTY does not apply to defects resulting from any action of the purchaser including but not limited to mishandling improper interfacing operation outside of design limits improper repair or unauthorized modification This WARRANTY is VOID if the unit shows evidence of having been tampered with or ...

Reviews: